IT·SCIENCE

New technology cuts blood amino acid analysis from 40 minutes to 6

by
Koo Bon-hyuk
Published : Nov. 26, 2026 - 10:41:38
    • Copy Completed!

View Korean Original

- KRISS establishes precise measurement method for 14 amino acids

- Advance expected to improve reliability of newborn metabolic disease screening

Researchers examine the mixed standard solution containing 14 amino acids developed by KRISS. [KRISS]
Researchers examine the mixed standard solution containing 14 amino acids developed by KRISS. [KRISS]

South Korean researchers have developed a technology capable of precisely analyzing 14 amino acids in blood in under six minutes.

The Korea Research Institute of Standards and Science (KRISS) announced Thursday that a research team led by Senior Researcher Jeong Ji-seon of the organic measurement group had developed a phosphate-buffered mixed standard solution containing 14 amino acids and begun distributing it in earnest.

Amino acid levels in blood are key indicators for assessing the body's metabolic and nutritional status and for early screening of metabolic diseases in newborns. However, errors can arise across the multiple steps involved from sample preparation to analysis, making a reliable measurement standard essential.

Conventional mixed amino acid standard solutions rely primarily on acidic solutions to improve the solubility and storage stability of their components. This made it difficult to include acid-sensitive amino acids such as tryptophan and glutamine in a single solution. The representative standard solution distributed by the US National Institute of Standards and Technology (NIST) is also acid-based.

The research team overcame this limitation by using a phosphate buffer that keeps acidity at a neutral level, successfully incorporating all 14 amino acids — including acid-sensitive components — into a single stable standard solution. The result is effectively a reference "answer key" that provides accurate concentration values for each component.

The team also established a reference measurement method that separates and quantifies all 14 amino acids in under six minutes without the need for derivatization, a separate chemical pretreatment step. The new approach significantly reduces the time required compared with conventional methods, which typically take more than 40 minutes, and also cuts down on error factors introduced during pretreatment.

Certified reference materials for clinical applications developed and distributed by KRISS. [KRISS]
Certified reference materials for clinical applications developed and distributed by KRISS. [KRISS]

The team also verified measurement accuracy in actual human blood samples, which have complex compositions.

Building on the development of this standard solution, KRISS has established a one-stop support system that allows diagnostic companies and hospital laboratories to manage everything — from instrument calibration and sample processing to verification of analytical results — in accordance with national standards.

Amino acids are the third area, after glucose and creatinine, in which KRISS provides one-stop support across the entire measurement process.

"The core achievement here is that we simplified the analytical procedure and incorporated components that existing products had missed into the standard solution," Senior Researcher Jeong said. "Going forward, we will establish national standards applying the same framework to areas such as female hormones and vitamins, and continue expanding our research scope."

The findings were published in the Journal of Chromatography A, an international academic journal in the field of analytical chemistry.